Thermo Scientific™ iQ Series Gas Analyzers provide access to analyzer data at any time.
Product Overview
With the Thermo Scientific™ iQ Series Gas Analyzer, people can access the analyzer's data, day or night. Designed to become more intelligent over time, the iQ Series Gas Analyzers reduce troubleshooting time by providing a tough yet highly accurate solution.
The Thermo Scientific™ 146iQ Multi-Gas Calibrator delivers the same high level of performance when calibrating ambient air and other gas monitoring instruments. The Multi-Gas Calibrator's output is used to calibrate instruments performing zero, precision, Level 1 span checks, audits, and multi-point routines, in accordance with U.S. EPA requirements.
Now backed by a two-year warranty, the iQ Series Gas Analyzer platform offers continuous intelligence through the following key features:
Predictive Diagnostics
Enhanced serviceability and plug-and-play components help resolve issues before they occur. Interchangeable DMC modules provide a robust and precise solution, while 24/7 access to analyzer data minimizes troubleshooting time.
Proactive Communications
Maintain performance with iQ360 and email alerts. When subscribed to iQ360, the instrument automatically notifies the technical support team upon alarm activation. The support team then prepares a corrective plan and sends it via email to designated staff. The iQ Series helps reduce surprises and keep operations running smoothly.
Personal Device Connectivity
The iQ companion mobile app enables real-time, continuous remote monitoring. Users can monitor their analyzers, receive maintenance alerts, access product documentation, and contact support directly through the app.
Specifications
Source: Thermo Scientific - Chemical Analysis Production Process and Analytics
|
Detection Limits |
1.5 ppb (60 sec. averaging time) |
Flow Rate |
0.8 L/min. |
Inputs |
16 Inputs (TLL), 4 Isolated Voltage Inputs 0 to 10 V |
Linearity |
±0.25 % full scale |
Response Time |
20 sec. (10 sec. averaging) (Photometer), <180 sec. (Ozonator Response Time) |
Span Drift |
<1 % full scale (1 month) |
Temperature (English) Operating |
32 °F to 113 °F |
Temperature (Metric) Operating |
0 °C to 45 °C |
Zero Noise |
0.75 ppb RMS (60 sec. averaging time) |
Accuracy |
±0.8 % of reading +0.2 % of full-scale |
Certifications/Compliance |
CE, TUV-SUD Safety, UKCA |
Communication Ports |
Serial ports: 1 RS-232/485 port 1 RS-485 external accessory port
Other ports: 3 Full Speed USB ports (one in front, two in rear) 1 Gigabit ethernet port
Communication: MODBUS, Streaming, Gesytec (Bayern-Hessen)
|
Depth (English) |
24 in. |
Depth (Metric) |
609 mm |
Description |
146iQ Multi-Gas Calibrator |
Electrical Requirements |
100-240 VAC 50/60 Hz, 275 watts |
Height (English) |
8.72 in. |
Height (Metric) |
221.48 mm |
Outputs |
6 Isolated voltages outputs 6 Isolated analog current outputs |
Precision |
1 ppb |
Repeatability of Flow Control |
±0.2 % of full scale |
Voltage |
100/240 V |
Weight (English) |
43 lb. (with 3 MFCs, Ozonator, and Photometer) |
Weight (Metric) |
19.5 kg (with 3 MFCs, Ozonator, and Photometer) |
Width (English) |
16.75 in. |
Width (Metric) |
425.45 mm |
Zero Drift |
<1.0 ppb (24 hours) |
Unit Size |